Meinhard nebulizer – an industrial standard

the MEINHARD nebulizers are standard nebulizers that are suitable for all major ICP instruments on the market and is used as standard by several instrument manufacturers.

Further information:
https://www.meinhard.com/

High sensitivity Typ A

Argon: 1 l/min
pressure: 30 psi - 50 psi
sample pickup: +        
aerosol stability: +

for samples with some solids, Typ C

Argon: 1 l/min
pressure: 30 psi - 50 psi
sample pickup: ++
aerosol stability: ++

Ar Low Flow, Typ K

Argon: 0,7 l/min
pressure: 30 psi - 50 psi
sample pickup: ++
aerosol stability: ++

Alle three versions are available with 0.5 / 1 / 2 / 3 / 4 or 5 ml/min sample throughput.
Both in glass or quartz.

HEN – nebulizer

The HEN nebulizer version is especially for small amounts of available sample. Slightly equal detection limit but up 25 times less sample needed. The HEN can be used with 100 µL/min and up to 170 psi. Common they are used with half the pressure  µL/min sample throughput. Fitting all standard spray chambers.

DIHEN – nebulizer

the pneumatic DIHEN nebulizer is used for direct injetion of smallest samples into ICP or ICP/MS withou tthe use of a spray chamber. Used for very rare or toxic samples.

For further information visit the Meinhard web page or send us an email.